Error Message

Problem

Solution

Locked Rotor?

B

B

(continued)

There may be an internal problem. For internal problems, call Customer

Service.

Lost PHASE SYNC!

There may be an internal problem.

1.

Turn the unit’s power off and
on, adjust the settings, and
attempt to resume pumping.

2.

If the problem persists, call
Customer Service.

Must SET RPM!

A

A

1.

Press the RPM key.

2.

Use the UP and DOWN
ARROW keys
to adjust the
RPM setting.

3.

Press the START/STOP key to
start the pump.

B

B
1.

Press the RPM key.

2.

Use the UP and DOWN
ARROW keys
to adjust the
RPM setting.

3.

Press the CALIBRATE key to
restart the calibration.

Must SET TIME!

1.

Press the TIME key.

2.

Toggle from SECONDS to
MINUTES to HOURS with the
TIME key.

3.

Use the UP and DOWN
ARROW keys
to adjust the
TIME setting.

4.

Press the START/STOP key to
start the pump.

.

Troubleshooting Guide

7

7-3

You have attempted to pump in
the RPM Mode with the RPM
set to 0 rpm.

You have attempted to dispense
fluid during calibration with the
RPM set to 0 rpm.

You have attempted to
pump in the TIME
Mode

with the TIME set

to 00:00:00.