Cdtx-90-1 calibrate menu – Omega CDTX-90 User Manual

Page 10

Advertising
background image

page 8

Select CUSTOM only if you are connecting a certifi ed conductivity sensor. Select STANDARD

for

all

other

systems.

For STANDARD sensors: Select from these options: 0.01, 0.1, 1.0, 10.0 or 20.0.

For CUSTOM sensors: Enter the precise cell constant from the certifi cate provided with your

sensor, or from the information label on the sensor.

Select from the following options: μS, mS, kΩ, MΩ,

PPM

If the Units selection is PPM, you can set the ratio of Total Dissolved Solids to μS. See page

15 for additional information. The TDS factor can only be set in PPM.

Adjust the temperature of the system based on an accurate external reference.

For use with STANDARD sensors: Perform this single-point wet calibration for most accurate

results. Not required for CUSTOM sensors except for periodic system accuracy confi rmation.

Enter all zeroes to restore factory calibration to TEMP and COND settings.

Select the input source to be associated with the current loop output: Conductivity or

Temperature.

Select the minimum and maximum values for the current loop output.

Be sure to modify this setting if you change the Cond. Units.

Select the source for the Open Collector output: Conductivity or Temperature

Select the mode of operation for the Open Collector output.

Options available are High, Low or proportional Pulse.

The signal may be disabled (Off) if not in use.

In Low or High Mode, this Open Collector output will be activated when the process reaches

this value. Be sure to modify this setting if you change the Cond. Units.

The Open Collector output will be deactivated at Setpoint ± Hysteresis, depending on High or

Low Setpoint selection. (See details in section 3.3.)

In Pulse mode, set the process values where the proportional pulse will start and where it will

reach the maximum rate. Be sure to modify this setting if you change the Cond. Units.

In Pulse mode, set the maximum rate for the proportional Pulse output. The CDTX-90 will

accept any value from 0 to 400.

Use this “note pad” to record important dates, such as annual recertifi cation or scheduled

maintenance.

CDTX-90-1 Calibrate Menu

Display

(Factory settings shown)

Description

Cell

Constant:

Standard >

Cell:

Standard

1 >

Cell: Custom

1.0000 >

Cond Units:

uS >

PPM Factor:

2.00 >

Set:

Temperature >

Set:

Conductivity >

Loop Source:

Cond >

Loop Range: uS

0.0000

→ 100.000

>

Output Source:

Cond >

Output Mode:

Low >

Output Setpnt:

10.0000

uS >

Output Hys:

0.5000

uS >

Output Rng: uS

10.0000

→ 40.0000

>

Output PlsRate:

120

Pulses/Min >

Last

CAL:

04-20-07 >