Troubleshooting guide – Omega 3200ST User Manual

Page 23

Advertising
background image

M-4271/0707, pg. 23 of 26

H.

Troubleshooting Guide

Symptom

Possible Cause

Method of Correction

Clicking noise from

controller

Unit in error mode

Check there is sufficient

pressure and that the flow
path is not restricted or
blocked

Unit wired incorrectly

Check wiring is according to
Section B5

Loose connection

Check all connectors and
wiring

Damaged connector pins

Contact Omega Engineering

Blocked flow path

Check flow path for
obstructions.

Piping leak before sensor

Check all piping and
connections.

Insufficient power

Check the power supply
output and increase if

necessary

Output load resistance too low

Ensure the voltmeter or
data acquisition system or

display has an impedance of
greater than of 2.5kohm

Flow too low for the unit

Ensure that the flow being
measured is within the
capabilities of the unit

No response

Unit damaged or faulty

Contact Omega Engineering

Particles in flow path

Add filtration before the
sensor.

Flow path obscured

Remove any debris or
blockage in the flow path
eg. PTFE tape.

Unit calibrated for a different

gas

Check calibration certificate

and apply a “K” Factor to
readings if necessary.

Gas composition is variable

Contact Omega Engineering

Fittings have been changed

Replace the factory installed
fittings

Moisture in gas

Ensure gas is clean and dry

Insufficient warm-up period

Allow the unit to warm-up
for at least 5 minutes.

Zero drift

Verify the zero and adjust
as necessary as explained in

Section C

The gain potentiometer has

been adjusted

Contact Omega Engineering

Unit needs recalibration

Contact Omega Engineering

Inaccurate control

Flow too high for the unit

Ensure that the flow being
measured is within the

capabilities of the unit