unit may be damaged or fail prematurely. Such damage will not be

repaired under warranty.

Units should be installed in a clean, dry environment with an ambient

temperature that is as stable as possible. Avoid areas with strong magnetic

fields, strong air flows or excessive vibration.


In order to operate the differential pressure across the controller should be

in the range 15-45psid (1-3 bar). For optimum performance a differential
pressure of 25psid is recommended.


For Example, consider the following system:

The differential pressure across the flow controller in this system would be

100 psi – 14 psi = 86 psid. Consequently the flow controller would NOT be
able to control flow. For the unit to operate at optimum performance the

supply pressure from the gas cylinder would need to be lowered to 39 psig

to give 39 psi -14 psi = 25 psid.