M-4271/0707, pg. 24 of 26

Symptom

Possible Cause

Method of Correction

Insufficient or varying power

Check the power supply
output and increase if
necessary

Insufficient pressure

Ensure the pressure (an

differential pressure) is high
enough o operate the unit.

Varying pressure

Check the stability of the
pressure regulation and
improve if necessary.

Ambient temperature too high
or too low

Place the unit in a suitable
environment

Output load resistance too low

Ensure the voltmeter or
data acquisition system or
display has an impedance of
greater than of 2.5kohm

Gas temperature too high or
too low

Ensure the gas temperature
is within the recommended
operating range

Inaccurate control

Unit damaged or faulty

Contact Omega Engineering

Gas flow through unit not
completely stopped

Ensure there is no flow
through the unit. The
easiest way to do this is to

plug both the inlet and
outlet.

Severe fluctuations in the
ambient temperature e.g. unit

in direct sunlight

Carry out the rezero
procedure in a stable

environment

Unstable power supply

Check the stability and
suitability of the power

source

Problems with
rezeroing

Insufficient warm-up period

Allow the unit to warm-up
for at least 5 minutes.