TABLE IV PRESSURE DROPS

TABLE V APPROXIMATE WEIGHTS

5.

OPERATING INSTRUCTIONS

5.1

Preparation and Warm Up

It is assumed that the Mass Flow Meter or Controller has been correctly installed
and thoroughly leak tested as described in section 2. Make sure the flow source
is OFF. Power up the transducer using your own power supply (or turn the
POWER switch to the ON position at the front panel of your FMA14PD Series
Command Module). Allow the Mass Flow Meter or Controller to warm-up for a
minimum of 15 minutes.

9

MODEL

WEIGHT

SHIPPING WEIGHT

FMA 1500 ranges up to 10 L/min transmitter

1.71 lbs (0.78 kg)

3.21 lbs (1.46 kg)

FMA 1500 ranges above 10 L/min transmitter

2.42 lbs (1.10 kg)

3.92 lbs (1.78 kg)

FMA 1400 ranges up to 10 L/min transmitter

2.20 lbs (1.00 kg)

3.70 lbs (1.68 kg)

FMA 1400 ranges above 10 L/min transmitter

2.84 lbs (1.29 kg)

4.34 lbs (1.97 kg)

MODEL

FLOW RATE

[std liters/min]

MAXIMUM PRESSURE DROP

[mm H

2

O]

[psid]

[mbar]

FMA 1500

up to 10

25

0.04

2.5

15

63

0.09

6.4

20

300

0.44

30

30

800

1.18

81

40

1480

2.18

150

50

2200

3.23

223

60

3100

4.56

314

80

4422

6.50

448

100

5500

8.08

557

FMA 1400

up to 10

720

1.06

75

15

2630

3.87

266

20

1360

2.00

138

30

2380

3.50

241

40

3740

5.50

379

50

5440

8.00

551

60

7480

11.00

758

80

10204

15.00

1034

100

12850

18.89

1302

Advertising