Omega FMA 1500 User Manual

Page 25

Advertising
background image

8.2

Troubleshooting Guide

20

REMEDY

check connection of power supply

disconnect FMA 1400/1500 transducer
from power supply; remove the short-
ing condition or check polarities; fuse
resets automatically

disconnect power cord from AC
supply; remove and inspect fuses at
AC power input connector of FMA14PD
Series; replace as necessary

REMOVE CAUSE OF SHORT CIRCUIT!

flush clean or disassemble to remove
impediments or replace

flush clean or disassemble to remove
impediments or return to factory for
replacement

return to factory for replacement

re-adjust valve (section 7.4)

disconnect FMA 1400/1500 transducer
from power supply; remove the short-
ing condition or check polarities; fuse
resets automatically

REMOVE CAUSE OF SHORT CIRCUIT!

apply appropriate gas pressure

flush clean or disassemble to remove
impediments or replace

signal and power supply commons are
different

apply appropriate gas pressure

check cables and all connections or
replace

re-adjust set point

re-adjust valve (section 7.4)

LIKELY REASON

power supply off

fuse blown
(FMA 1400/1500)

fuse blown
(FMA14PD Series)

filter screen obstructed at inlet

occluded sensor tube

pc board defect

valve adjustment wrong

fuse blown
(FMA 1400/1500)

inadequate gas pressure

filter screen obstructed at inlet

ground loop

inadequate gas pressure

cable or connector
malfunction

set point is too low
(<2% of full scale)

valve adjustment wrong

INDICATION

lack of
reading or
output

output reads
at (+) or (- )
saturation
only

flow reading
does not
coincide with
the set point
(FMA 1400
models only)

no response
to set point
(FMA 1400
models only)