TABLE VI FMA 1400 SOLENOID VALVE ORIFICE SELECTION TABLE

8.

TROUBLESHOOTING

8.1

Common Conditions

Your Mass Flow Meter/Controller was thoroughly checked at numerous quality
control points during and after manufacturing and assembly operations. It was cal-
ibrated in accordance to your desired flow and pressure conditions for a given gas
or a mixture of gases.

It was carefully packed to prevent damage during shipment. Should you feel that
the instrument is not functioning properly please check for the following common
conditions first:

Are all cables connected correctly?

Are there any leaks in the installation?

Is the power supply correctly selected according to requirements?
When several meters are used a power supply with appropriate
current rating should be selected.

Were the connector pinouts matched properly?
When interchanging with other manufacturers' equipment, cables
and connectors must be carefully wired for correct pin configurations.

Is the pressure differential across the instrument sufficient?

19

ORIFICE PART NUMBER

FLOW RATE [N2]

OR.010

under 10 sccm

OR.020

10 to 1000 sccm

OR.040

1 to 5 slpm

OR.055

5 to 10 slpm

OR.063

10 to 15 slpm

OR.073

15 to 20 slpm

OR.094

20 to 50 slpm

OR.125

50 to 100 slpm