Teledyne Refrigerated Sampler 6712FR User Manual

Page 152

Advertising
background image

6712FR Refrigerated Sampler
Section 7 Remote Operation

7-10

Disable

This command is similar in functionality to a pin F low signal (a signal typically
used to enable or disable the sampler from a flow meter). At run time, the sam-
pler starts with the phone enable in the enabled state. PHONE DISABLE is
valid while a program is running. If successful, a reply of REMOTE ENABLE IS
FALSE will be returned.

A Remote Disable is the highest priority disable. If the sample is disabled with
this command, no condition other than command 6#, ENABLE, will allow the
program to become enabled again.

Enable

This command is used to re-enable a phone-disabled sampler. PHONE
ENABLE is valid while a program is running. If successful, a reply of REMOTE
ENABLE IS TRUE will be returned.

Grab Sample

This command causes the sampler currently running a program to take a sam-
ple. The sample will count as one of the program’s samples and is placed in
the current bottle (Part A’s current bottle when using two-part programming).
TAKE_SAMPLE is valid while the program is running, after the start time, but
not while in the MANUAL PAUSE screens. If successful, a response similar to
PUMPING XXML SAMPLE INTO BOTTLE ## will be returned. If not, a
response similar to NO SAMPLE, PROGRAM WAITING START will be
returned.

Table 7-4 6712 Remote Phone Commands (Continued)

Phone
Command

Name

Description

5

#

6

#

7

#

Advertising