Teledyne Refrigerated Sampler 6712FR User Manual

Page 78

Advertising
background image

6712FR Refrigerated Sampler
Section 4 Standard Programming

4-22

Figures 4-5, 4-6, 4-7, 4-8, and 4-9 show examples of reports con-
figured to show program settings, sampling results, combined
results, and a summary report.

(The Source and Error Codes Listed below refer to why a sample was taken and what happened while it was being
taken.)

Source Codes

*C

Calibrate sample

Sample volume delivered during calibration.

*D

Disable

Sample event at disable time.

*E

Enable

Sample event at enable time.

Ev

Event

Event-paced sample event.

*F

Flow

Flow-paced sample event.

*G

Grab sample

Sample volume delivered as a grab sample.

M

Command driven sample A command driven sample.

PH

Phone command

Phone command sample.

R

Resume

Sample event at resume time.

*S

Start

Sample event at start time.

Sw

Switch time

Sample event at switch time.

TO

Time Override

Sample event at bottle-switch time.

*T

time

Time-paced sample event.

Error Codes

*DJ

Distributor Jammed

The distributor jammed.

*IP

sample in progress

Sample in progress when report was printing.

*L

pump latch open

Unable to take sample because the pump band was opened.

*ND no distributor arm

There is no distributor arm or the stop is damaged.

*NL

No Liquid Detected

The sampler was unable to detect liquid.

LF

Not enough flow

Not enough flow for the minimum sample volume entered for samples with volumes
proportional to flow.

*NM No More Liquid

After the sampler detected liquid and while the sample was being taken, the liquid
detector stopped detecting liquid.

M

Bottle Full

The composite bottle is full.

*O

Sampler Shut Off

The sampler was turned off with the On-Off key during the sample event.

Ov

Probable Overflow

The sampler was directed to take a sample that would not fit into the bottle. Occurs
for flow paced, time override programs only.

*P

Power Failed

The power supply failed during the sample event. Note: Power failures during five
consecutive sample events results in a LOW BATTERY shut-down.

*PJ

Pump Jammed

The pump jammed.

*Sk

Sample Skipped

Sample not taken.

*US User Stopped

You pressed the Stop key during the sample event.

Table 4-3 Reports: Program Events, Source Codes, and Error Codes (Continued)

Advertising