6712FR Refrigerated Sampler
Section 4 Standard Programming

4-32

8

PROGRAM MODULE?

YES NO

Select YES to program the module.

9

MODE OF OPERATION:

FLOWMETER

LEVEL AND VELOCITY

The 750 can have two different modes of operation: Flowmeter or Level and
Velocity.
For this example, choose Flowmeter. Press

.

10

FLOW CALCULATION:

AREA * VELOCITY

LEVEL TO FLOW RATE

When you select Flowmeter, you will have two options for performing the
flow conversion: Area * Velocity or Level to Flow Rate.
For this example, choose Area * Velocity. Press

.

11

LEVEL TO AREA:

CHANNEL SHAPE

DATA POINTS

For this example, choose Channel Shape. Press

.

12

ROUND PIPE

U-CHANNEL

RECTANGULAR CHANNEL

TRAPEZOIDAL CHANNEL

For the channel shape, select Round Pipe. Press

.

13

ROUND PIPE

DIAMETER = 2.00 ft

Enter the diameter of the round pipe. Press

.

14

CURRENT LEVEL IS

__.__ ft

ADJUST LEVEL TO

__.__ ft

The current level will be displayed. Enter an adjustment if needed. If the
adjustment is different by more than 0.50 ft, you will be prompted “

ARE

YOU

SURE

?

YES

NO

” Choose

NO

to re-enter; select

YES

if correct. Press

.

15

ZERO LEVEL OFFSET:

_.___ ft

Enter the amount of zero level offset. Press

.

16

MAXIMUM DEPTH FOR

VELOCITY MEASUREMENT

2” 3” 4”

(This screen is applicable for standard-sized probes only.)

Select the maximum depth for the velocity measurement. Press

.

17

NUMBER OF BOTTLES:

1 2 4 8 12 24

Select the number of bottles in your bottle kit by pressing either arrow key
until the correct number blinks. Press

.

From this point forward, the screens will be the same as in Table 4-2.