Output (Fat Monitor Mode) Format

a Transmission Data

a Control Commands

Remote Mode

All types of data may be entered by connecting the BF-350 to a PC and switching to Remote mode.

When in Remote mode, the body fat analyzer may be operated remotely by sending control commands from a host.

a Error Codes

If an error occurs, the body composition analyzer will send an error code to the host.

a Switching to Remote Mode

The receiving PC must be ready to accept output data immediately after
measurement is completed.

ƽ

CAUTION

Output data

Byte Length

First byte

1

1

5

Second byte

K: unit of weight kg

L: unit of weight lb

S: stability of load (weight output)

D: instability of load (weight output)

I : overload status

---: negative weight output

Header

Weight

xxx.x

Output data format

¥ Terminal data are CR (ASCII format: 0DH), LF (ASCII format: 0AH).
¥ Measurement data will be sent in the format shown to the right:

x x x x x . x

1. When the unit is switched off, press the [

]

and [

] keys simultaneously.

ƽ

Note

The default setting is off.

2. Press [0], [8].

3. Switch Remote mode on and off by pressing [1]

or [0].

[1] : Enables Remote mode
[0] : Disables Remote mode

4. Press [

] to confirm.

Only "r S" will appear (standby screen).

ƽ

Note

To switch off Remote mode, start
from the beginning again and
select "OFF" at Step 3.

ATHLETIC

STEP ON

FEMALE

MALE

kg

cm / ft.in

STABILIZED

AGE

% FAT

P T

ATHLETIC

STEP ON

FEMALE

MALE

kg

cm / ft.in

STABILIZED

AGE

% FAT

P T

ATHLETIC

STEP ON

FEMALE

MALE

kg

cm / ft.in

STABILIZED

AGE

% FAT

P T

ATHLETIC

STEP ON

FEMALE

MALE

kg

cm / ft.in

STABILIZED

AGE

% FAT

P T

*All fixed-length data right aligned. Enter

for blanks.

0

U? : Confirm current setting

U0: Use cm/kg mode

U1: Use in/lb mode

LF

CR

1

U

LF

CR

0

U

LF

CR

?

U

commands

U

R? : Confirm current setting

R0: Select equation for Westerners

R1: Select equation for Orientals

LF

CR

1

R

LF

CR

0

R

LF

CR

?

R

commands

R

D? : Confirm current setting

D0 : Set clothing weight (5-byte fixed length)

D1 : Set gender male = 1 female = 2

D2 : Set body type standard = 0 athlete = 2

D3 : Set height (5-byte fixed length)

D4 : Set age (2-byte fixed length)

D6 : Set Target BF (2-byte fixed length)

LF

CR

X

X

6

D

LF

CR

X

X

4

D

LF

CR

X

X

X

X

X

3

D

LF

CR

X

2

D

LF

CR

X

1

D

LF

CR

X

.

X

X

X

0

D

LF

CR

?

D

commands

D

G1: Start measuring body fat

G2: Measure weight only

LF

CR

2

G

LF

CR

1

G

commands

G

CHR$(01FH): Reset settings

LF

CR

CHR$(01FH)

Special

commands

E00 : Attempted to start measuring without completing settings

E01 : Error in calculating body fat percentage

EXX :other error ’ body fat analyzer switches off

LF

CR

X

X

E

LF

CR

1

0

E

LF

CR

0

0

E

*See p.26 "7. Troubleshooting" for details of errors.