Product specifications – Tanita Body Composition Analyzer BC-418MA User Manual

Page 21

Advertising
background image

GB

GB

Product specifications

Output Items

Measurement System

Measurement Frequency

Impedance

Measurement

Weight

Measurement

Input Items

8 polar Bioelectrical Impedance Analysis

50kHz

Measurement Current

90µA

Electrode Material

Pressure Contact Stainless Steel Foot Pads, The Hand Grips are Metal-Plated to ABS Resin.

Measurement Style

Whole body, Right leg, Left leg, Right arm and Left arm

Measurement Range

150 ~ 1200

Measurement System

Maximum

Capacity

Minimum

Graduation

Strain Gauge Load Cell

Accuracy at first calibration

200 kg / 0.1 kg

±0.2kg

Clothes Weight

0-200 kg / 9 st 13.8 lb / 440lb

0.1 kg increments

Gender

Male / Female

Body Type

Standard / Athletic

Age

7~99 years old /1year increments

Height

Display

Print-out

Display

90 - 249 cm

1 cm increments

Target Body Fat %

4~55%

Weight

0 - 200 kg

Target Body Fat %

4~55%

Gender

Male / Female

Body Type

Standard / Athletic

Age

7 - 99 years old / 1 year increments

Height

90 - 249 cm

1 cm increments

FAT%

1 - 75% / 0.1% increments

Body Type

Standard / Athletic

Gender

Male / Female

Age

7 - 99 years old / 1 year increments

Height

90 - 249 cm

1 cm increments

Weight

2 - 200 kg

0.1 kg increments

BMI

0.1 increments

BMR

1 kJ increments / 1 kcal increments

Impedance 1

150 ~ 1200

/ 1

increments

FAT% 2

1 - 75% / 0.1% increments

Fat Mass 2

0.1 kg increments

FFM 2

0.1 kg increments

TBW

0.1 kg increments

Predicted Weight

0.1 kg increments

Predicted fat mass

0.1 kg increments

FAT TO LOSE/GAIN

0.1 kg increments

Predicted muscle Mass 2

0.1 kg increments

Others

Desirable Range for FAT% and FM (Standard and 20-79 years old ONLY)

3 Rows, 5 Digits LCD

Output Data Interface

RS-232C (D-sub 9 pins Female Connector)

Temperature Range of Usage

0ºC/35ºC

Power Consumption

Relative Humidity

17.5 W

BF-418MA

Accuracy Grade

MDD : CLASS 2a

Power source

AC adaptor (included) Center Minus MODEL SA25-0535U CLASS 2

Input Voltage: 100-240 VAC 50/60 Hz 550mA Output Voltage: 5 VDC ± 0.3 V

Rated Current: 3.5 A No Load Input Voltage: 5 VDC ± 0.3 V

MDD : CLASS 3

30% / 80% (without condensation)

Weight of Equipment

12kg

Size

Accuracy at first calibration

12

Weighing Platform

340 X 375 X 90mm

830mm

Height

/

1 This item displays Whole Body, Right Leg, Left Leg, Right Arm and Left Arm.
2 This item displays Right Leg, Left Leg, Right Arm, Left Arm and Trunk.

The current BMR retrogression formula is a formula which acts on the principle of using the FFM value
from the results of body composition measurement according to the BIA, A good relationship is shown in
the BMR value based on actual breath analysis REE or R=0.9 (p<0.0001). These results were presented at the
First Annual Nutrition Week (American College of Nutrition, American Society for Clinical Nutrition,
American Society for Parenteral and Enteral Nutrition, North American Association for the Study of
Obesity) held in 2002 in San Diego.

NOTE: This model has been calibrated for those between ages of 18-84.

Those individuals outside of this age range may not be obtain accurate readings.

As shown in Figure 1: the REE (BMR) has a stronger relationship to the FFM than to body weight, and a
difference is visible between males and females in the distribution trends. We see that in principle that we
should calculate from the FFM rather than by the old formula centered on the relationship with weight.

<Fig: 2> Comparison of BMR Values from the TANITA Multiple regression model and Breath Analysis

(Presented at Nutrition Week Held in San Diego in 2002)

Advertising