Appendix D: Performance Verification

DĆ8

MTG100 & MTG300 MPEG Generator User Manual

MPEG generator rear panel

75 W BNC cable

Oscilloscope (TDS784D)

75 W to 50 W

minimum loss attenuator

Figure DĆ5: Initial equipment connection for verifying the SSI output

2. Set the oscilloscope controls as indicated below (only the horizontal axis

setting is different from the setup in the ASI Output test):

Displayed channel

CH1

. . . . .

Vertical axis

200 mV/div

. . . . . . . . . .

Horizontal axis

10 ns/div

. . . . . . .

Horizontal position

Center

. . . .

Record length

500

. . . . . . . .

Acquire mode

Sample

. . . . . . . .

Acquire sequence

RUN/STOP button only

. . . . .

Trigger mode

Auto

. . . . . . . . .

Trigger level

0 V

. . . . . . . . .

Trigger source

CH1

. . . . . . . .

Trigger position

50 %

. . . . . . .

Trigger slope

Rising Edge

. . . . . . . . .

Trigger coupling

DC

. . . . . .

Display style

Dots

. . . . . . . . .

Input coupling

DC

. . . . . . . .

Input impedance

50

W

. . . . . .

Measure

Amplitude

. . . . . . . . . . . . .

3. Open the test40.trp file on the MPEG generator.

a. Select Open TS File... from the FILE pull-down menu.

b. In the resulting Select TS File dialog box, select the test40.trp file, and

then press the OK bezel button.

4. Press the START/STOP button on the MPEG generator to start the signal

output of the test40.trp file.

5. Verify that the following LEDs on the MPEG generator’s front panel are

illuminated: PLL, BUFFER, and HDD.

6. Use the oscilloscope to measure that the signal amplitude is within the range

of 720 mV to 880 mV.