Tutorials

2Ć24

MTG100 & MTG300 MPEG Generator User Manual

Figure 2Ć20: Insertion Setup dialog box

5. Use the arrow buttons to highlight the PID parameter in the Insertion Setup

dialog box as shown in Figure 2-20.

6. Use the rotary knob to set the PID value to 18 (0x12), and then press the

Apply bezel button to accept the settings and close the dialog box.

7. Press the START/STOP button to start transport stream output. When

transport stream output is started, the Output Status indicator in the status
bar changes from Stopped to Running.

8. Press the MENU button to enable the Icon cursor in the hierarchy display.

9. Use the rotary knob to move the Icon cursor to the bottom of the hierarchy

display.

Observe that an INS_SI icon is added at the end of the hierarchy as shown in
Figure 2-21. The icon’s background is colored blue to indicate it has been
added by the user.

10. Connect a 75

W cable between the ASI OUT connector on the MPEG

generator rear panel to the ASI IN connector of the MPEG test system.

11. Set up the MPEG test system to monitor the transport stream output from the

MPEG generator. Figure 2-22 shows the transport stream data displayed on a
Tektronix MTS200 Series Real Time Analyzer with the GHOST icon shown.
In this case, the inserted EIT appears as a GHOST, since it is not referenced
in an appropriate table in the transport stream.