Front Panel Controls

Front Panel LCD

23063L-003 Rev. 2 12/23/04

R110Xi/R170Xi User Guide

49

INIT FLASH MEM

Initialize Flash Memory

Caution • Perform this operation only when it is necessary to erase all
previously stored information from Flash memory.

1.

When the LCD displays

INITIALIZE CARD

, press the right oval to select

YES

. (If you are prompted for a password, enter your password using the

instructions in

Password-Protected Parameters on page 43

.)

The LCD asks

INITIALIZE FLASH

.

2.

Press the right oval to select

YES

.

The LCD asks

ARE YOU SURE?

.

3.

Do you wish to continue?

• Press the left oval to select

NO

to cancel the request and return to the

INIT

FLASH MEM

prompt.

• Press the right oval to select

YES

and begin initialization.

When formatting is complete,

INIT FLASH MEM

displays.

SENSOR PROFILE

Sensor Profile
The media sensor profile may be used to troubleshoot registration problems that
may be caused when the media sensor detects preprinted areas on the media or
experiences difficulty in determining web location.
• Press the right oval to start this standard calibration procedure and print a media

sensor profile.

Figure 31

shows a media sensor profile. If the sensitivity of the media and/or ribbon

sensors must be adjusted, use

Calibrate Media and Ribbon Sensors on page 50

to

adjust the media and ribbon sensor sensitivity.

Figure 31 • Media Sensor Profile