Technical specifications, B. test b - two minute tightness test – RIDGID FG-100 User Manual

Page 9

Advertising
background image

IMPORTANT

The leak test B countdowns are programmable from keyboard. Read the

Instruction manual included in the CD-ROM.

Connect the pressure input

to the leak test kit (cat. 26028).

Connect the kit to the gas network.

Pressurize the network at the testing pressure.

Close the valve on "T" connector.

Select the “START” option pressing:

Press

to start the let-by test procedure.

Wait 1 minute instrument countdown for network stabilization.
Wait 2 minute instrument countdown for leak measurement.

Print the report.

P1

If required, use the rubber blanking

plugs for connecting.

Open the valve on "T" connector.

Press:

8

Technical Specifications

This product is fully compliant with part 2 and part 3 of EN 50379

Type: 2 or 3 cells hand-held flue gas analyser/Boiler tuning unit.

Working temperature: -5°C to +45 °C (up to 50°C for short time)
Storage temperature: –20 to +50°C (3 months maximum at temperatures
exceeding the operational limits).
Dimensions: 115x70x300 mm
Weight: Nett. 1.1 kg battery and printer included

Internal Printer: Impact type 24 columns
Autozero: Automatic 60 sec. procedure at startup for sensor calibration
and instrument self diagnosis.
Printer power supply: using the analyser battery pack.
Power supply: High capacity Li-ion rechargeable battery pack / external
battery charger (90..264V 50/60Hz).
Charging time : approx 3h at 90% with instrument off.
Battery life: 7h continuous operation printer and back-light off).
Display: Large (40x56 mm), monocrome, graphic LCD display with
automatic back-light device.
Fuel types: Up to 10 totally programmable (the standard set depends by
the national local regulation).
Data memory: up to 250 complete analysis data structured by Tags.
Serial interface: USB standard interface.
Battery storage temperature: 0 to 35°C

Parameter

Range

Resolution

Accuracy

O

CO

CO diluted

NO

Tair

Tgas

Pressure

Draft

2

0 - 25%

0.1%

±0.1% vol.

0 - 8000ppm

1ppm

±10ppm <300ppm

±4%rdg up to 2000ppm

±10%rdg elsewhere

0.8 - 10%

0.05%

0 - 4000ppm

1ppm

±5ppm <125ppm

±4%rdg up to 4000ppm

-10 - 100°C

1°C

±(0.2% + 0.15°C)

0 - 1000°C

1°C

±(0.3% + 0.3°C)

0 - 200 mbar

0.1mbar

±0.5mbar <50 mbar

±1%rdg >50 mbar

±40.00 hPa

0.01 hPa

±0.03 hPa < 300 Pa

±1% rdg. > 300 Pa

H compensated

2

All emission measurements can be displayed with reference to a programmable O value.
Accuracy limits are stated as % of reading. An additional ±1 digit error has to be considered.
The stated pressure relative accuracy is valid only after the zero procedure.
Measuring reading can be directly converted from °C to °F, ppm to mg/Nm and from hPa to
mmH O, mbar, inH O.

2

2

2

3

B. Test B - two minute tightness test

Select and confirm the “

” option.

Test type

Change and confirm test type to “Test B” for two minute let-by test.

Press the “ZERO” function (F1) to zeroing the pressure sensor.

Press

to start the let-by test procedure.

Wait 1 minute instrument countdown for leak measurement.

Print the report. Press: