Calibration procedure – Activeforever Phoenix Model 1850 Knee CPM User Manual

Page 20

Advertising
background image

CALIBRATION PROCEDURE

MODEL 1850


TOOLS REQUIRED:

Straight Edge, Protractor, 3/32 Allen Wrench, Phillips Screwdriver



INSTRUCTIONS:

Turn power off
Remove knee potentiometer cover
Remove top cover from the unit


Move SW1 slide switch located at the bottom left of the board from run to calibrate

Using the up and down arrows move the unit to the zero position

Use a straight edge to determine zero

Look for these values on the hand control

Ext. window (Letter F with a red dot)

ROM Window (2) Flexion Window (490) reading this as 2490 (2490 represents the zero setting)


If these values are not present then the unit is out of calibration

If the above values do not appear on the hand control then it is necessary to move the knee pot to gain these values. To move
the knee pot, loosen the clamps on the knee pot and turn the knee pot clockwise or counterclockwise to achieve the above
readings. Once these values have been achieved then tighten the knee pot clamps.


To lock in the zero (0) position, depress the ext. button 1 time and listen for an audible beep.

By depressing the UP arrow button you can move the unit into the 90 degree (90) position, using a protractor or similar device,
stopping at this position

Once in the 90 degree position (90) and if you are in the approximate range a red dot will be visible in the flexion window.
If the red dot is visible then depress the flexion button 1 time and listen for an audible beep.

Switch unit from calibration mode to run mode
Set hand control at -5 extension, 110 Flexion, 0 pause, and 10 speed


Depress start/stop switch on the hand control and allow unit to travel through full range of motion. If the unit travels the full
range of motion without binding then the unit has been calibrated successfully.