CALIBRATION PROCEDURE

MODEL DC2480

TOOLS REQUIRED:

Straight Edge, Protractor, 3/32 Allen Wrench, Phillips Screwdriver

INSTRUCTIONS:

Turn power off

Remove knee potentiometer cover

Remove top cover from the unit

Move SW1 slide switch located at the bottom left of the board from run to calibrate

Using the up and down arrows move the unit to the zero position

Use a straight edge to determine zero

Look for these values on the hand control

Ext. window (Letter F with a red dot)

ROM Window (2) Flexion Window (795) reading this as 2795 (2795 represents the zero setting)

If these values are not present then the unit is out of calibration

If the above values do not appear on the hand control then it is necessary to move the knee pot to gain these values. To move

the knee pot, loosen the clamps on the knee pot and turn the knee pot clockwise or counterclockwise to achieve the above

readings. Once these values have been achieved then tighten the knee pot clamps.

To lock in the zero (0) position, depress the ext. button 1 time and listen for an audible beep.

By depressing the UP arrow button you can move the unit into the 90 degree (90) position, using a protractor or similar devic

e,

stopping at this position

Once in the 90 degree position (90) and if you are in the approximate range a red dot will be visible in the flexion window.

If the red dot is visible then depress the flexion button 1 time and listen for an audible beep.

Switch unit from calibration mode to run mode

Set hand control at

–10 extension, 120 Flexion, 0 pause, and 10 speed

Depress start/stop switch on the hand control and allow unit to travel through full range of motion. If the unit travels the

full

range of motion without binding then the unit has been calibrated successfully.