Replacing heel loop – Activeforever Invacare Tracer IV Wheelchair User Manual

Page 30

Advertising
background image

SECTION 4—FRONT RIGGINGS

Tracer

®

IV

30

Part No. 1110558

Replacing Heel Loop

NOTE: For this procedure, refer to FIGURE 4.6 on page 31.

1. Pull the cam lock lever up to unlocked position.

2. Perform one of the following as shown in Detail "A".

• Footplates with Spring Buttons: Push in the release buttons and remove the

footplate assembly from the front rigging support.

• Bolt-In-Place Footplates:

i. Using a screw driver to hold the threaded rivet in position, remove the button

head screw from the threaded rivet.

ii. Remove the threaded rivet and button head screw securing the footplate

assembly to the front rigging support.

3. Remove the mounting screw, spacer and locknut that secure the heel loop to the footplate.

4. Remove existing heel loop from slide tube.

5. Install new heel loop onto slide tube.

6. Install the mounting screw, spacer and locknut to secure the heel loop to the footplate.

Tighten until the spacer is secure.

7. Insert the lower footrest assembly into the upper footrest assembly to desired height.

8. Perform one of the following as shown in Detail "A":

• Footplates With Spring Buttons: Ensure that the release buttons fully protrude

from holes on both sides of the upper footrest support.

• Bolt-In-Place Footplates:

i. From the outside of the swingaway front rigging, insert the threaded rivet

through both the front rigging support and the footplate assembly.

ii. From the inside of the swingaway front rigging, insert the button head screw

through the appropriate adjustment hole and thread into the threaded rivet.

iii. Using a screwdriver to hold the threaded rivet in position, securely tighten the

button head screw. Torque to 32 in-lbs.

9. Rotate cam lock lever down to locked position.