13

EverGo User Manual

Getting Started

To start your EverGo device, follow these steps. (To

learn about the touch screen displays, turn to

the next section of this manual.)

1. Connect a nasal cannula to the oxygen

outlet on top of your device, as shown in

Figure 7.

Note: A standard single lumen nasal

cannula and tubing (not supplied) should

be used to deliver oxygen from the EverGo device. The device works with

cannula tubing up to 30 ft. (9 m).

2. To turn on the device, gently

touch the screen shown in

Figure 8. This first touch brings

the screen to life and shows the

on/off symbol, battery status, hour

reading and software version.

3. Gently touch the on/off symbol

to turn on the device.

This “double touch” feature helps

prevent unwanted starts or inadvertent changes in oxygen flow.

You may touch anywhere on the screen as your first touch.
The second touch must be on the on/off symbol

within

5 seconds or the screen will turn off.

You may begin breathing from your unit immediately after starting

the device.

VER 1.0.0.0

Figure 8. Touch the on/off symbol to

start. The number on the screen (i.e. Hr

= 421) shows the number of hours this

unit has been operated.

Figure 7. The nasal cannula is

connected to the oxygen outlet port

on top.

www.ActiveForever.com