3 short

beeps

+

1 long beep

No Breath Alarm
This alarm occurs when a breath

is not detected for a period of 2

minutes (120 seconds) or more�

This alarm can be silenced by

pressing the Alarm symbol

(the bell symbol) shown on the

screen� This alarm becomes silent

as soon as a breath is detected�

If no breath is detected after

approximately 30 minutes, the

unit shuts down to conserve

power� (This alarm is inactive for

ten minutes after startup�)

Check the

connection from

the cannula to

the device� Make

sure the nasal

cannula is properly

positioned on

your face and that

you are breathing

through your

nose� Make sure

the cannula tubing

is not kinked or

obstructed�

3 long beeps

+

1 short beep

Oxygen Concentration Alarm
This alarm occurs when the

device is delivering a lower

concentration of oxygen than

specified� This alarm can be

silenced by pressing the Alarm

symbol (the bell symbol shown

on the screen)�

Change to another

source of oxygen

and contact

your equipment

provider�