Ix71/ix51 – Olympus IX71 User Manual

Page 33

Advertising
background image

29

IX71/IX51

Fig. 42

}In general, the potential resolving power of an objective is fully utilized if the

diaphragm is stopped down to correspond with the numerical aperture
(N.A.) of the objective.

}Depending on the specimen, image contras or focal depth in observation

or photomicrography may be improved by keeping the aperture iris
diaphragm stopped down a little.
In general. a good image is obtained if the diaphragm is stopped down
to between 70% and 80% of the N.A. of the objective. Stop further down
for less contrasty specimens.

}To check the position of the perimeter of the aperture iris diaphragm,

remove the eyepieces and look into the eyepiece sleeves to view the
aperture iris diaphragm image and the objective’s exit pupil.

}When using the U-BI90CT binocular observation tube, the aperture iris

diaphragm can be observed by setting the turret to position “CT” or using
the U-CT30 centering telescope.

3

Flipping Up the Condenser Holder

(Fig. 43)

}To create working space and facilitate specimen replacement,

micromanipulator positioning or mounting an objective from the side of
the stage, flip up the condenser holder upwards.

1. Placing a hand against the bottom of the condenser holder, press it

upwards.

2. The tension of the flipping-up operation can be adjusted by turning the

adjustment screw @ using a coin.

# Adjust the tension so that the flipped-up condenser will not drop

down by itself.

# When returning a flipped-up condenser to its original position, do so

gently and slowly. Optimum illumination performance cannot be
obtained if the condenser is not positioned correctly.

2

Using the Aperture Iris Diaphragm

(Fig. 42)

Fig. 43

@

Objective exit pupil

70-80%

30-20%

Advertising
This manual is related to the following products: