Olympus IX71 User Manual

Page 38

Advertising
background image

34

4. Hold a phase contrast ring and, while pushing the spring ƒ inside the

turret with the edge of the phase contrast ring …, insert the ring com-
pletely in the turret position until the ring frame contacts the bottom of the
position.

(Fig. 48)

# Be careful not to apply pressure to the ring slit inside the frame.
5. Rotate the optical element centering knobs clockwise to tighten the

centering screws lightly.

# If the optical element centering knobs are mounted, the turret is un-

able to be rotated.

# Do not tighten the optical element centering knobs too much, for

this may deform the frames of the optical elements.

6. Place the index † provided with each optical element in the index inser-

tion hole having the same number Š as the number of the position ‡ in
which the corresponding optical element is mounted. (Figs. 49 & 50)

}Note that the IX-ULWCD does not have the index.
}The optical element with the number indicated by marking · ‰ is

engaged in the light path. (Fig. 49)

}To remove an optical element index, use the tip of a ball-point pen or

mechanical pencil.

7. When all of the required optical elements have been mounted, attach

the cover and tighten the detaching screws.