6

Setup

6

.1 Setup steps

A

fter mounting and electrical connection, VEGABAR 17 is ready for

operation.

VEGABAR

17

delivers a current of 4 … 20 mA corresponding to the

actual process pressure.

F

urther settings are not necessary.

6

.2 Recalibration

W

ith both instruments with thread ring or field housing, zero and span

can be adjusted via integrated potentiometers. Adjustment range:

l

Z

ero ± 5 %

l

S

pan ± 5 %

T

his allows, for example, the consideration of an installation position

different from the reference installation position.

A

shifting of zero shifts span also respectively.

Note:
T

he potentiometer for span should only be used if you have adequate

calibration equipment (at least 3 times more precise than the deviation
of VEGABAR 17).

R

ecommended recalibration cycle: 1 year.

P

roceed as follows:

1

L

oosen the plug connector and screw the screwed ring in

connected status

2

P

lace the plug connector onto the instrument place and pull both

carefully out of the instrument

Instruments with plug
connector or cable out-
let

18

VEGABAR

17