Nominal measuring ranges and overload resistance

Nominal range

Overload, max. pressure

Overload, min. pressure

G

auge pressure

0 … 0.1

bar/0 … 10 kPa

15

bar/1500 kPa

-1

bar/-100 kPa

0 … 0.2

bar/0 … 20 kPa

20

bar/2000 kPa

-1

bar/-100 kPa

0 … 0.4

bar/0 … 40 kPa

30

bar/3000 kPa

-1

bar/-100 kPa

0 … 1

bar/0 … 100 kPa

35

bar/3500 kPa

-1

bar/-100 kPa

0 … 2.5

bar/0 … 250 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

0 … 5

bar/0 … 500 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

0 … 10

bar/0 … 1000 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

0 … 25

bar/0 … 2500 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

-1 … 0

bar/-100 … 0 kPa

35

bar/3500 kPa

-1

bar/-100 kPa

-1 … 1.5

bar/-100 … 150 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

-1 … 5

bar/-100 … 500 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

-1 … 10

bar/-100 … 1000 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

-1 … 25

bar/-100 … 2500 kPa

50

bar/5000 kPa

-1

bar/-100 kPa

-0.05 … 0.05

bar/-5 … 5 kPa

15

bar/1500 kPa

-1

bar/-100 kPa

-0.1 … 0.1

bar/-10 … 10 kPa

20

bar/2000 kPa

-1

bar/-100 kPa

-0.2 … 0.2

bar/-20 … 20 kPa

30

bar/3000 kPa

-1

bar/-100 kPa

-0.5 … 0.5

bar/-50 … 50 kPa

35

bar/3500 kPa

-1

bar/-100 kPa

A

bsolute pressure

0 … 1

bar/0 … 100 kPa

35

bar/3500 kPa

0 … 2.5

bar/0 … 250 kPa

50

bar/5000 kPa

0 … 5

bar/0 … 500 kPa

50

bar/5000 kPa

0 … 10

bar/0 … 1000 kPa

50

bar/5000 kPa

0 … 25

bar/0 … 2500 kPa

50

bar/5000 kPa

Reference conditions and influencing variables (similar to DIN EN 60770-1)
R

eference conditions according to DIN EN 61298-1

-

T

emperature

18 … 30 °C (64 … 86 °F)

-

R

elative humidity

45 … 75 %

-

A

ir pressure

860 … 1060

mbar/86 … 106 kPa

(12.5 … 15.4

psi)

D

etermination of characteristics

limit point adjustment according to DIN 16086

S

upplement

VEGABAR

75 - 4 … 20

mA/HART