Diagnostics

Check the internal connections:
• Remove the housing cover.
• Loosen the three screws with a screwdriver and remove the oscillator

from the housing.

• Measure with an ohmmeter (range M

W) the resistance values be-

tween the following contacts:

Contact 4 to middle pin (1)

The resistance must be 1 M

W.

If the resistance is less, it means there is moisture in the housing or a
failure of the electrode insulation. A possible reason could be a
noninsulated electrode which is used in conductive (moist) product.
If the resistance is higher or if the connection is interrupted, the reason
is usually a bonding failure in the adapter plate or a resistor defective
due to strong electrostatic discharge.
In both cases, the measuring probe must be repaired by VEGA.