3 wiring plan – VEGA VEGAWELL 72 4 … 20 mA_HART User Manual
Page 20
1
U
nscrew the housing cover
2
I
nsert the connection cable through the cable entry into the
plastic housing
3
L
oosen the screws with a screwdriver
4
I
nsert the wire ends into the open terminals according to
the wiring plan
5
T
ighten the screws with a screwdriver
6
C
heck the hold of the wires in the terminals by lightly
pulling on them
7
T
ighten the compression nut of the cable entry. The seal
ring must completely encircle the cable
8
R
etighten the housing cover
T
he electrical connection is finished.
5
.3 Wiring plan
5
1
2
3
4
Fig. 7: Wire assignment, connection cable
1
brown (+): to power supply or to the processing system
2
blue (-): to power supply or to the processing system
3
yellow: is only required with VEGADIS 12, otherwise connect to minus
4
)
4
Shielding
5
Breather capillaries with filter element
4)
F
or customer-specific versions, already connected with blue (-) when
shipped.
Direct connection
20
VEGAWELL
72 • 4 … 20
mA/HART
5 C
onnecting to power supply
27630
-EN
-080806