VEGA VEGAWELL 72 4 … 20 mA_HART User Manual

Page 36

Advertising
background image

Nominal range

Overload, max. pressure Overload, min. pressure

A

bsolute pressure

0 … 15

psi

500

psi

0

psi

0 … 35

psi

700

psi

0

psi

0 … 70

psi

900

psi

0

psi

0 … 150

psi

1300

psi

0

psi

0 … 350

psi

1900

psi

0

psi

Output variable

O

utput signal

4 … 20

mA/HART

F

ailure signal

22

mA (3.6 mA), adjustable via PACTware

M

ax. output current

22.5

mA

R

un-up time approx.

2.5

s

D

amping (63 % of the input variable)

0 … 10

s, adjustable

S

tep response time

200

ms (ti: 0 s, 0 … 63 %)

F

ulfilled NAMUR recommendations

NE

43

Reference conditions and actuating variables (similar to DIN EN 60770-1)

R

eference conditions according to DIN EN 61298-1

-

T

emperature

+

15 … +25 °C (+59 … +77 °F)

-

R

elative humidity

45 … 75 %

-

A

ir pressure

860 … 1060

mbar/86 … 106 kPa

(12.5 … 15.4

psig)

D

etermination of characteristics

L

imit point adjustment according to

IEC

61298-2

C

haracteristics curve

linear

R

eference installation position

upright, diaphragm points downward

I

nfluence of the installation position

<

0.2

mbar/20 Pa (0.003 psig)

Deviation determined according to the limit point method according to IEC 60770

7

)

A

pplies to digital HART interface as well as to analogue current output 4 … 20 mA.

S

pecifications refer to the set span. Turn down (TD) is the relation nominal measuring

range/set span.

7)

I

ncl. non-linearity, hysteresis and non-repeatability.

36

VEGAWELL

72 • 4 … 20

mA/HART

10 S

upplement

27630

-EN

-080806

Advertising