7 maintenance and fault rectification, 1 maintenance, 2 remove interferences – VEGA VEGADIS 12 User Manual

Page 20: 7 maintenance and fault rectification

Advertising
background image

7 Maintenance and fault rectification

7.1 Maintenance

When used as directed in normal operation, VEGADIS 12 is

completely maintenance free.

7.2 Remove interferences

The operator of the system is responsible for taken suitable

measures to remove interferences.

VEGADIS 12 offers maximum reliability. Nevertheless faults

can occur during operation. These may be caused by the

following, e.g.:

l

Sensor

l

Process

l

Supply

l

Signal processing

The first measure to take is to check the output signal as well

as the atmospheric pressure compensation. The procedure is

described below. Further comprehensive diagnostics can be

carried out on a PC with the software PACTware™ and the

suitable DTM. In many cases, the causes can be determined in

this way and faults can be rectified.

However, if these measures are not successful, call the VEGA

service hotline in urgent cases under the phone no. +49 1805

858550.
The hotline is available to you 7 days a week round-the-clock.

Since we offer this service world-wide, the support is only

available in the English language. The service is free of

charge, only the standard telephone costs will be charged.

First of all open the housing cover. The indicated measured

value must not change. However, if the indicated value

changes nevertheless, the compensation of the atmospheric

pressure is not ensured. Check the breather facility on the

housing and the capillaries in the special cable.

Connect a handheld multimeter in the suitable measuring

range according to the wiring plan.

Reaction in case of failures

Causes of malfunction

Fault rectification

24 hour service hotline

Check pressure compensa-

tion

Checking the 4 … 20 mA sig-

nal

20

VEGADIS 12

Maintenance and fault rectification

20591

-EN
-071128

Advertising