Setup
In the window "Integration time“ you can enter
a time of max. 10 seconds. Then click to
OK
.
Sensor parameter adjustment –
Outputs
Click in the window "Instrument data param-
eter adjustment“ to
Outputs
.
Click in the window "Conditioning“ to
Current
output
.
In the window "Current output“ you can set
the strength of the current in case of failure
(e.g. line break). Furthermore, the current
output can be inverted, i.e. at 0 % the output
current is 20 mA, at 100 % 4 mA. Click to
Save
when you have modified the adjust-
ment.
Click in the window "Conditioning“ to
Display
of measured value
(only possible if you
have connected a display on VEGABAR).
In the window "Sensor-Display“ you choose
the parameter and the unit which the display
should indicate. If you choose "scaled“, you
can enter the figures corresponding to 0 %
and 100 %. Click to
Save
when you have
finished the adjustments in this window.