VEGA VEGABAR 65 4 … 20 mA_HART User Manual
Page 58
D
eviation
-
T
urn down 1 : 1 up to 5 : 1
<
0.075 %
-
T
urn down > 5 : 1
<
0.015 %
x TD
D
eviation with absolute pressure measuring range 0.1 bar
-
T
urn down 1 : 1 up to 5 : 1
<
0.25 %
-
T
urn down > 5 : 1
<
0.05 %
x TD
Influence of the product or ambient temperature
A
pplies to instruments with digital signal output (HART, Profibus PA, Foundation Fieldbus) as well
as to instruments with analogue current output 4 … 20 mA. Specifications refer to the set span.
T
urn down (TD) = nominal measuring range/set span.
T
hermal change zero signal and output span, reference temperature 20 °C (68 °F):
-
I
n the compensated temperature range
0 … +100 °C (+32 … +212 °F)
< (0.05 % + 0.1 %
x TD)
-
O
utside the compensated temperature
range
< (0.05 % + 0.15 %
x TD)
A
pplies also to instruments with analogue 4 … 20 mA current output and refers to the set span.
T
hermal change, current output
<
0.15 %
at -40 … +80 °C (-40 … +176 °F)
Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)
A
pplies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well as to analogue
current output 4 … 20 mA. Specifications refer to the set span. Turn down (TD) = nominal
measuring range/set span.
L
ong-term drift of the zero signal
< (0.1 %
x TD)/year
Ambient conditions
A
mbient, storage and transport temperature
-40 … +80 °C (-40 … +176 °F)
Process conditions
T
he specifications to the pressure stage and the product temperature are used as an overview.
T
he specifications of the type label are applicable.
P
ressure stage, process fitting
-
T
hread 316L
PN
60
-
T
hread Aluminium
PN
25
-
H
ygienic fittings 316L
PN
6, PN 10, PN 25, PN 40
-
F
lange 316L
PN
16/PN 40
or 150 lb/300 lb
P
roduct temperature
-
S
tandard
-12 … +140 °C (+10 … +284 °F)
-
with extenation, extended thread or
C
lamp 2½"
-12 … +140 °C (+10 … +284 °F)
58
VEGABAR
65 • 4 … 20
mA/HART
10 S
upplement
27531
-EN
-081209