Electrical connection

VEGAPULS 54K Ex with pressure-tight encapsulated connection
compartment in four-wire technology

• Four-wire technology, supply and output signal via two separate two-wire cables for opera-

tion in Ex-Zone 1 (VEGAPULS …K Ex) or Ex-Zone 0 (VEGAPULS …K Ex0).

• Output signal 4 … 20 mA (active).
• Optional external indicating instrument with analogue and digital display (can be mounted

up to 25 m away from the sensor in Ex area).

• Adjustment with PC, HART

®

handheld or adjustment module MINICOM (can be plugged into

the sensor or into the external indicating instrument VEGADIS 50).

• Load max. 500

Ω.

2

-

+

4 … 20 mA

2

2

2

4

1)

250

EEx d ia

1)

If the resistance of the processing systems
connected to the 4 … 20 mA signal output is less
than 200

Ω, a resistor must be connected to the

connection cable during adjustment to get a loop
resistance between 250

Ω up to 300 Ω.

The digital adjustment signal would otherwise be
severely damped or short-circuited due to
insufficient resistance of the connected process-
ing system. Digital communication with the PC or
the HART

®

handheld would not be ensured.

VEGA-
CONNECT 2

VEGADIS 50 Ex

Ex area

active

2)

HART

®

handheld

Non Ex area

2)

4 … 20 mA active means that the sensor delivers
a level-dependent current of 4 … 20 mA (source).
The measuring signal of the sensor reacts
electrically to the processing system (e.g. display)
like a current source.