4 adjustment with hart® handheld, 4 adjustment with hart, Handheld – VEGA VEGAFLEX 54K User Manual

Page 31: The most important adjustment steps

Advertising
background image

VEGAFLEX 54K

31

The most important adjustment steps

You will find the complete adjustment proce-
dure via HART

®

protocol in the operating

instructions of the HART

®

handheld.

To enter parameters, first press "

ENTER".

The entered value will be saved in the
handheld, but not in the sensor itself.
Press "

SEND" to transmit the entered value to

the sensor.

Press "

OK" to acknowledge the following

warning. Follow the instructions on the dis-
play.

Press again "

OK", and the adjustment that

was just carried out is displayed.

Generic: SENSOR

1 LRV

5.850 m

2 PV URV

0.300 m

HELP

HOME

Generic: SENSOR
PV LRV

5.850 m

0.300

m

HELP DEL ABORT ENTER

Generic: SENSOR

1 LRV

5.850 m

2 PV URV

0.300 m

HELP SEND HOME

250 ... 350

+

-

SPS

Setup

4.4 Adjustment with HART

®

handheld

The VEGAFLEX sensors can be set up with
any HART

®

handheld. A special DDD (Data

Device Description) is not necessary. Just
connect the HART

®

handheld to the signal

cable after having connected the sensor to
power supply.

Note

If the resistance of the signal current circuit is
less than 200

, a resistor of 250 … 350

must be connected to the signal/connection
cable during adjustment. Simply loop the
resistor into the sensor connection cable
(see figure).

PLC