VEGAPULS 51P … 54P

24890-EN-041227

Set-up

Adjustment with medium

Fill the vessel e.g. to 10 % and enter 10 % in
the menu "

Min. adjust“ with the "+“ and "–“

keys. Then fill the vessel, e.g. to 80 % or
100 % and enter 80 % or 100 % in the menu
"

Max. adjust“ with the "+“ and "–“ keys.

percentage value and then the product dis-
tance corresponding to the percentage value.

Note:

The difference between the adjustment val-
ues of the lower product distance and the
upper product distance should be as large
as possible, preferably at 0 % and 100 %. If
the values are very close together, e.g. lower
product distance at 40 % (3.102 m) and
upper product distance at 45 % (3.331 m),
the measurement will be less accurate. A
characteristic curve is generated from the
two points. Even the smallest deviations
between actual product distance and en-
tered product distance will considerably
influence the slope of the characteristic
curve. If the adjustment points are too close
together, small errors inflate to considerably
larger ones when the 0 % or the 100 % value
is outputted.

OK

+

or

100.0%

at

m (d)

XX.XXX

The distance indication flashes
and you can choose "feet“ and
"m“.

Confirm the adjustment with
"

OK“.

With "

+“ and "–“ you adjust the

percentage value for the min.
value (example 0.0 %).

The entered percentage value
is written in the sensor and the
min. distance value corres-
ponding to that percentage
value flashes.

With the "

+“ or "–“ key you can

assign a level distance (ex-
ample 5.85 m) to the previ-
ously adjusted percentage
value. If you do not know the
distance, you have to do a
sounding.

The adjusted product dis-
tance is written in the sensor
and the display stops flash-
ing.

You thereby adjusted the lower product dis-
tance as well as the percentage filling value
corresponding to the lower product distance.

Note:

For level detection outside the operating range,
the operating range must be corrected accord-
ingly in the menu "

Sensor optimisation/Operat-

ing range“.

+

OK

Ad-
just-
ment
in

m(d)

0.0%

at

m (d)

XX.XXX

OK

+

or

(max. adjustment)

Now you enter the max. adjustment (upper
product distance) (example 100 % and 1.270
m product distance). First you have to enter the

Signal
condit
ioning

Scal
ing

0 %
corres
ponds

XXXX

100 %
corres
ponds

XXXX

Deci-
mal
point

888.8

prop.
to

Mass

Unit

Kg

with
medium

Min.
adjust
at %

XXX.X

Max.
adjust
at %

XXX.X

Advertising