Settings in the "Extended menu“

7.7 Simulation

7.8 Special function: Reset

To check the outputs and connected instru-
ments, you can adjust in this menu item any
individual percentage value with the keys [+]
and [–]. The starting point is always the ac-
tual measured value. The indicated value
flashes when the simulation is activated. 15
minutes after setting the simulated value, the
simulation is automatically terminated and the
display reset to measured value indication.

Param.
TAG-
No. 1

Adjust-
ment

Signal
condit-
ioning

Out-
puts

Simu-
lation

Simu-
lation
%

XX,X

Simu-
lation
Now!
OK?

TAG -
No. 1
%
xx,x

Spe-
cial
func-
tions

Reset
Level

Reset

OK ?

Reset
Now!

OK ?

Param.
TAG-
No. 1

Adjust-
ment

Signal
condit-
ioning

Out-
puts

Simu-
lation

TAG -
No. 1
%
xx,x

With this reset, all parameter values of the
measurement loop TAG-No. 1 are reset to
factory settings (empty and full adjustment,
relay outputs, current output, volt output and
scaling of the meas. value indication are
reset).