VEGA VEGASON 65 4 … 20 mA_HART four-wire User Manual
Page 17
5
P
lug the connection head to the transducer tube. Do not squeeze
any cables. The cylinder is in the correct position, if the lower
notch (2) of the transducer tube is no longer visible
1
2
3
4
Fig. 8: Mounting - Version D
1
Groove for locking the connection head
2
Notch
3
Hexagon nut SW 46
4
Terminal screws, connection head (hexagon spanner - size 4)
6
T
ighten the screws (4) on the cylinder moderately with an Allen
wrench (size 4)
7
R
otate the connection head in such a way that the cable entry
points downwards (moisture can drain off). Fasten the cylinder
with screws (4); torque max. 5 Nm
8
I
nsert the plug at the end of the transducer cable into the socket on
the remote electronics
1
2
3
Fig. 9: Plug connector between transducer and socket of electronics housing
1
Socket electronics housing
2
Plug connector
3
Connection piece, transducer tube
VEGASON
65 • 4 … 20
mA/HART four-wire
17
4 M
ounting
28782
-
EN
-081127