5 switch-on phase – VEGA VEGASON 63 Profibus PA User Manual

Page 22

Advertising
background image

22

5 Connecting to power supply

VEGASON 63 • Profibus PA

28786-EN-130417

1

2

3

Display

1 2

I2C

Fig. 21: Connection compartment, double chamber housing
1 Spring-loaded terminals for voltage supply

2 Plug connector for VEGACONNECT (I²C interface)

3 Ground terminal for connection of the cable screen

I2C

1

1

2

Fig. 22: Wiring plan, double chamber housing
1 Voltage supply, signal output

5.5 Switch-on phase

After VEGASON 63 is connected to voltage supply or after voltage

recurrence, the instrument carries out a self-check for approx. 30

seconds. The following steps are carried out:

Internal check of the electronics

Indication of the instrument type, the firmware as well as the sen-

sor TAGs (sensor designation)

Status byte goes briefly to fault value

Then the current measured value will be displayed and the corre-

sponding digital output signal will be outputted to the cable.

1)

Connection compartment

Wiring plan

Switch-on phase

1)

The values correspond to the actual measured level as well as to the settings

already carried out, e.g. default setting.