Process temperature (transducer tem-

perature)

-40 … +80 °C (-40 … +176 °F)

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

8)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug

9)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap M20 x 1.5; 1 x blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

Ʋ Double chamber housing

– 1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5; 1 x blind plug M16 x 1.5 or

optionally available with 1 x plug M12 x 1 for external

display and adjustment unit

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT, 1 x blind

plug M16 x 1.5 or optionally 1 x plug M12 x 1 for exter-

nal display and adjustment unit

or:

– 1 x plug (depending on the version), 1 x blind plug

M20 x 1.5; 1 x blind plug M16 x 1.5 or optionally avail-

able with 1 x plug M12 x 1 for external display and

adjustment unit

Spring-loaded terminals for wire cross-

section

< 2.5 mm² (AWG 14)

Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry

Ʋ Single chamber housing

1 x IP 68 cable gland M20 x 1.5; 1 x blind plug M20 x 1.5

Ʋ Double chamber housing

1␣x IP␣68 cable gland M20␣x␣1.5; 1␣x blind stopper

M20␣x␣1.5; 1␣x blind stopper M16␣x␣1.5

Connection cable

Ʋ Wire cross-section

0.5 mm² (AWG 20)

Ʋ Wire resistance

< 0.036 Ω/m (0.011 Ω/ft)

Ʋ Tensile strength

< 1200 N (270 lbf)

Ʋ Standard length

5 m (16.4 ft)

Ʋ Max. length

1000 m (3280 ft)

Ʋ Min. bending radius

25 mm (0.984 in) with 25 °C (77 °F)