11 Secure screw (3) with lacquer or similar to ensure permanent
tightness of the swivelling holder
12 Rotate the connection head in such a way that the cable entry
points downwards (moisture can drain off). Fasten the cylinder
with screws (5); torque max. 5 Nm
13 Insert the plug at the end of the transducer cable into the socket on
the remote electronics
1
2
3
Fig. 7: Plug connector between transducer and socket of electronics housing
1
Socket electronics housing
2
Plug connector
3
Connection piece, transducer tube
Version D is supplied in three parts:
l
Transducer
l
Transducer cable
l
electronics for wall mounting
Assemble the version as follows:
1
Loosen the hexagon nut (3) on the transducer cable
2
Insert the transducer tube from below into the mounting opening
G
1A
3
Fasten (SW 46) the hexagon nut (3)
4
Remove the plug from below out of the connection head and plug
it into the socket of the transducer tube
Mounting - Version D
16
VEGASON
64 • Profibus PA
4 Mounting
28787
-EN
-081127