3 shorten suspension cable, 4 instrument repair – VEGA VEGAWELL 51 4 … 20 mA User Manual

Page 21

Advertising
background image

?

4 … 20

mA signal missing

l

W

rong connection to power supply

à Check connection according to chapter "Connection

steps" and if necessary, correct according to chapter
"Wiring plan"

l

N

o voltage supply

à Check cables for breaks; repair if necessary

l

supply voltage too low or load resistance too high

à Check, adapt if necessary

I

n Ex applications, the regulations for the wiring of intrinsically

safe circuits must be observed.

D

epending on the failure reason and measures taken, the

steps described in chapter "Set up" must be carried out again,
if necessary.

6

.3 Shorten suspension cable

S

horten the suspension cable individually. Proceed as follows:

1

R

emove the filter adapter from the transparent capillary line

2

C

ut the suspension cable with an edge cutter to the

requested length

Caution:
D

o not squeeze the capillary cable, this will influence the

pressure compensation. If necessary, rework with a sharp
knife.

3

R

emove approx. 10 cm of the cable mantle, strip off

approx. 1 cm of the wire ends
I

nsert the filter adapter

T

he work steps are finished.

6

.4 Instrument repair

I

f a repair is necessary, please proceed as follows:

Y

ou can download a return form (23 KB) from our Internet

homepage www.vega.com under: "Downloads - Forms and
certificates - Repair form
".

B

y doing this you help us carry out the repair quickly and

without having to call back for needed information.

Reaction after fault rec-
tification

VEGAWELL

51 • 4 … 20

mA

21

M

aintenance and fault rectification

32465

-

EN

-080604

Advertising