VEGA VEGAWELL 51 4 … 20 mA User Manual

Page 26

Advertising
background image

S

ignal resolution

2 µA

F

ailure signal

22

mA

M

ax. output current

22

mA

R

un-up time

2

s

S

tep response time

100

ms (ti: 0 s, 0 … 63 %)

F

ulfilled NAMUR recommenda-

tions

NE

43

Reference conditions and actuating variables (similar to DIN EN 60770-1)

R

eference conditions according to DIN EN 61298-1

-

T

emperature

+

15 … +25 °C (+59 … +77 °F)

-

R

elative humidity

45 … 75 %

-

A

ir pressure

860 … 1060

mbar/86 … 106 kPa

(12.5 … 15.4

psig)

D

etermination of characteristics

L

imit point adjustment according to

IEC

61298-2

C

haracteristics curve

linear

R

eference installation position

upright, diaphragm points downward

I

nfluence of the installation position

<

0.2

mbar/20 Pa (0.003 psi)

Deviation determined according to the limit point method according to IEC 60770

4

)

S

pecifications refer to the set span. Turn down (TD) = nominal measuring range/set

span.

D

eviation

-

T

urn down 1 : 1 up to 5 : 1

<

0.2 %

-

T

urn down up to 10 : 1

<

0.04 %

x TD

Influence of the product or ambient temperature
S

pecifications refer to the set span. Turn down (TD) = nominal measuring range/set

span.

Average temperature coefficient of the zero signal

I

n the compensated temperature range of 0 … +80 °C (+32 … +176 °F), reference

temperature 20 °C (68 °F).

4)

I

ncl. non-linearity, hysteresis and non-repeatability.

26

VEGAWELL

51 • 4 … 20

mA

S

upplement

32465

-EN

-080604

Advertising