2 Connect the black ground cable (6) to the ground screw (5)
3 Plug the green connection cables to the cable gland (4)

and the green connection cables to the thermostat (7)

together

4 Plug the red connection cables to PLICSCOM (2) and the

red connection cables to the thermostat (3) together

5 Insert the plug connector into the excavation in front of the

oscillator

6 Place the indicating and adjustment module PLICSCOM

(1) in the desired position on the electronics (you can

choose any one of four different positions - each displaced

by 90°)

7 Press PLICSCOM (1) lightly onto the electronics and turn it

to the right until it snaps in.

Fig. 4: Installation of PLICSCOM

Note:

If you intend to retrofit the sensor with a PLICSCOM for

continuous measured value indication, a higher cover with an

inspection glass is required.

Heating for the indicating and adjustment module PLICSCOM

11

Connecting to voltage supply

31708