3 Product description

3.1 Configuration

If the sensor is already ordered with a heating, then the heating

is already mounted.
However, the heating can be mounted also afterwards.

The scope of delivery encompasses:

l

Indicating and adjustment module PLICSCOM with heating

mounted in the sensor

l

Documentation

- a device operating instructions manual

- this supplementary instructions manual

The scope of delivery encompasses:

l

Indicating and adjustment module PLICSCOM with heating

l

Thermostat

l

Cable gland with plug M12x1

l

Documentation

- this supplementary instructions manual

The heating must not be mounted in instruments with Ex

approval.

3.2 Principle of operation

A heating can be integrated so that data can be read out on the

indicating and adjustment module even at extremely low

temperatures.
The heating must be powered with a protective extra-low

voltage according to VDE 0106 part 101. For further

information see also in chapter "Technical data".
The heating for the indicating and adjustment module

PLICSCOM is suitable for all plics

®

sensors with single

chamber housing and the external indicating and adjustment

unit VEGADIS 61.
The heating is permanently prepared and must not be

activated in addition. It switches on automatically with

temperatures around -5 °C (+23 °F) and switches off again at

approx. 0 °C (+32 °F).

Components

Scope of delivery - heating

already mounted

Scope of delivery - retrofitting

set

Area of application

Heating for the indicating and adjustment module PLICSCOM