VEGA VEGAFLEX 66 (-110…+400°C) Foundation Fieldbus User Manual
Page 48
Output variable
Output
-
Signal
digital output signal, Foundation Fieldbus
protocol
-
Physical layer
according to IEC 61158-2
Channel Numbers
-
Channel 1
Primary value
-
Channel 2
Secondary value 1
-
Channel 3
Secondary value 2
Transmission rate
31.25 Kbit/s
Current value
10 mA, ±0.5 mA
Accuracy (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
-
Temperature
+
18 … +30 °C (+64 … +86 °F)
-
Relative humidity
45 … 75 %
-
Air pressure
860 … 1060 mbar/86 … 106 kPa
(
12.5 … 15.4 psig)
Deviation in characteristics and characteristics
Reference installation conditions
-
Flange size
> DN
100
-
min. distance to installations
>
500 mm (19.69 in)
-
Vessel diameter
1 m (3.281 ft) centrical installaton
-
Medium
Water
Temperature drift (current output)
0.06 %/10 K relating to the max. measuring
range
Accuracy
see diagrams
48
VEGAFLEX
66 • Foundation Fieldbus
10 Supplement
32765
-EN
-080716