VEGA VEGAFLEX 62 Foundation Fieldbus User Manual

Page 18

Advertising
background image

Make sure that the probe is not subjected to strong lateral forces.
Mount VEGAFLEX 62 at a position in the vessel where no
disturbances, e.g. from filling openings, agitators, etc., can occur.

Fig. 11: Lateral load

If there is a danger of the probe touching the vessel wall during
operation due to product movements or agitators etc., the measuring
probe should be securely fixed.

In the gravity weight there is a thread (M12), e.g. for a ring bolt
(optional) - (article no. 2.27424).

Make sure that the probe cable is not completely taut. Avoid tensile
loads on the cable.

Avoid undefined cable-vessel connections, i.e. the connection must be
either grounded reliably or isolated reliably. Any deviation from this
requirement can lead to measurement errors.

Inflowing medium

Fixing

18

VEGAFLEX

62 • Foundation Fieldbus

4 Mounting

31840

-EN

-100426

Advertising