VEGA VEGAFLEX 62 Foundation Fieldbus User Manual

Page 46

Advertising
background image

-

Trimming accuracy - rod

< 1

mm (0.039 in)

-

Cable: ø 6 mm (0.236 in)

up to 60 m (196.9 ft)

-

Trimming accuracy - cable

±0

.05 %

Lateral load with rod: ø 16 mm (0.63 in)

30 Nm (22 lbf ft)

Max. tensile load with cable: ø 6 mm
(0.236 in)

30 KN (6745 lbf)

Thread in gravity weight (cable version)

M

12

Input variable

Measured variable

Level of liquids and solids

Min. dielectric figure of the medium

ε

r

> 1

.6

2

4

5

3

1

2

4

5

3

1

Fig. 25: Measuring ranges of VEGAFLEX 62

1

Reference plane

2

Probe length L

3

Measuring range (default setting refers to the measuring range in water)

4

Upper dead band (see diagrams under Accuracy - grey section)

5

Lower dead band (see diagrams under Accuracy - grey section)

Output variable

Output

-

Signal

digital output signal, Foundation Fieldbus protocol

-

Physical layer

according to IEC 61158-2

46

VEGAFLEX

62 • Foundation Fieldbus

10 Supplement

31840

-EN

-100426

Advertising