7 Maintenance and fault rectification

VEGACAP 27 • - Relay (DPDT)

33758-EN-130916

6. Loosen the two screws with a Torx screwdriver.
7. Open the tensile ring on the old electronics module and pull out

the old electronics module.

8. Compare the new electronics module with the old one. The type

label of the electronics module must correspond to that of the old

electronics module.

9. Insert the new electronics module.
10. Screw in and tighten the two screws with a Torx screwdriver.
11. Lead the connection cable through the cable gland.
12. Insert the wire ends into the open terminals according to the wir-

ing plan and tighten them.

13. Check the hold of the wires in the terminals by lightly pulling on

them.

14. Tighten cable gland and check on tightness. The seal ring must

completely encircle the cable.

15. Carry out a fresh adjustment. See chapter "Set up, new adjust-

ment".

16. Screw the housing cover back on
The electronics exchange is now finished.

7.4 How to proceed in case of repair

You can find a repair form as well as detailed information on how to

proceed under www.vega.com/downloads and "Forms and certifi-

cates".
By doing this you help us carry out the repair quickly and without hav-

ing to call back for needed information.
If a repair is necessary, please proceed as follows:

Print and fill out one form per instrument

Clean the instrument and pack it damage-proof

Attach the completed form and, if need be, also a safety data

sheet outside on the packaging

Please contact for the return shipment the agency serving you. You

can find the agency on our home page www.vega.com.