VEGA VEGAFLEX 66 (-200…+400°C) 4 … 20 mA_HART two-wire User Manual

Page 16

Advertising
background image

E

xcessive system vibration or shocks, e.g. caused by agitators or

turbulence in the vessel (e.g. from fluidisation) can cause the coax
probe of VEGAFLEX 66 to vibrate in resonance. Should a coax probe
of more than 1 m (3.281 in) length should be used, you can provide a
suitable support or guy directly above the end of the probe to stabilise
it.

I

f there is a danger of the probe touching the vessel wall during

operation due to product movements or agitators etc., the measuring
probe should be securely fixed.

I

n the gravity weight there is a thread (M12), e.g. for a ring bolt

(

optional) - (article no. 2.27424).

M

ake sure that the probe cable is not completely taut. Avoid tensile

loads on the cable.

A

void undefined cable-vessel connections, i.e. the connection must be

either grounded reliably or isolated reliably. Any deviation from this
requirement can lead to measurement errors.

VEGAFLEX

66

should be integrated in the vessel insulation. This

prevents the electronics from strong heating or cooling by temperature
radiation.

M

ake sure that with heated or cooled vessels the permissible ambient

temperature on the housing is not exceeded. The permissible ambient
temperature is stated in chapter "Technical data" under "Ambient
conditions
".

Fixing

Temperature insulation

16

VEGAFLEX

66 • 4 … 20

mA/HART two-wire

4 M

ounting

34173

-EN

-100427