3

.2 Principle of operation

VEGAFLEX

66

is a level sensor with coax, rod or cable probe for

continuous level measurement in products with temperatures between
-

200 °C

and +400 °C (-328 … +752 °F).

I

t is designed for industrial use in all areas of process technology and

can be used in liquids.

H

igh frequency microwave pulses are guided along a steel rope or a

rod. Upon reaching the product surface, the microwave pulses are
reflected. The running time is evaluated by the instrument and
outputted as distance.

T

wo-wire electronics 4 … 20 mA/HART for power supply and

measured value transmission over the same cable.

T

he supply voltage range can differ depending on the instrument

version.

T

he data for power supply are specified in chapter "Technical data".

T

he optional background lighting of the indicating and adjustment

module is powered by the sensor. A certain level of operating voltage
is required for this. You can find the exact voltage specifications in
chapter "Technical data".

T

he optional heating requires its own operating voltage. You can find

details in the supplementary instructions manual "Heating for indicat-
ing and adjustment module
".

T

his function is generally not available for approved instruments.

3

.3 Operation

VEGAFLEX

66

can be adjusted with different adjustment media:

l

with indicating and adjustment module

l

with the suitable VEGA DTM in conjunction with an adjustment
software according to the FDT/DTM standard, e.g. PACTware and
PC

l

with manufacturer-specific adjustment programs AMS™ or PDM

l

W

ith a HART handheld

3

.4 Packaging, transport and storage

Y

our instrument was protected by packaging during transport. Its

capacity to handle normal loads during transport is assured by a test
according to DIN EN 24180.

T

he packaging of standard instruments consists of environment-

friendly, recyclable cardboard. For special versions, PE foam or PE foil
is also used. Dispose of the packaging material via specialised
recycling companies.